Project

General

Profile

Actions

Bug #2151

closed

Feature #1066: Development and Unit Testing

Module #1281: Bug correction

Activity #1740: Resolve issues reported by KDISC

Issue in sensitivity and specificity calculation

Added by Thara S Pillai over 2 years ago. Updated about 1 year ago.

Status:
Closed
Priority:
Normal
Assignee:
Target version:
Start date:
02/08/2024
Due date:
02/08/2024
% Done:

100%

Estimated time:
Planned Due Date:

Description

Issue report from client

We have checked 89 images from 14.01.2019 and observed that when we manually calculated, we obtained a sensitivity of 90% and a specificity of 100%.

The web portal shows a sensitivity value of 91.30% and a specificity value of 100%.

There is a discrepancy in sensitivity values.
Excel calculations and a screenshot of the web portal are attached for your reference.

Kindly request you to check and let us know how the sensitivity has arrived at 91.30% on the web portal.


Files

Actions #2

Updated by Thara S Pillai over 2 years ago

update status

Actions #3

Updated by Thara S Pillai over 2 years ago

  • Assignee changed from Shameera S L to Sreeshna T
Actions #4

Updated by Thara S Pillai over 2 years ago

Write DB function for returning the data as per the excel and compare with client shared excel data

Actions #5

Updated by Thara S Pillai over 2 years ago

Connect to SDC co-located server and check to see records from eye unidentifiable state to identify the issue.

Actions #6

Updated by Sreeshna T over 2 years ago

  • Status changed from New to Feedback
  • % Done changed from 0 to 100

count issue because of grade is good and error observation is eyetype unidentifiable considered as good in true positive case.

Actions #7

Updated by Sreeshna T over 2 years ago

found 6 patients whose gradability is bad and error observation is EyeTypeUnidentifiable from IDRSTL DB.

Actions #8

Updated by Thara S Pillai about 1 year ago

  • Status changed from Feedback to Closed
Actions