Bug #2151
closedFeature #1066: Development and Unit Testing
Module #1281: Bug correction
Activity #1740: Resolve issues reported by KDISC
Issue in sensitivity and specificity calculation
100%
Description
Issue report from client
We have checked 89 images from 14.01.2019 and observed that when we manually calculated, we obtained a sensitivity of 90% and a specificity of 100%.
The web portal shows a sensitivity value of 91.30% and a specificity value of 100%.
There is a discrepancy in sensitivity values.
Excel calculations and a screenshot of the web portal are attached for your reference.
Kindly request you to check and let us know how the sensitivity has arrived at 91.30% on the web portal.
Files
Updated by Thara S Pillai over 2 years ago
- Assignee changed from Shameera S L to Sreeshna T
Updated by Thara S Pillai over 2 years ago
Write DB function for returning the data as per the excel and compare with client shared excel data
Updated by Thara S Pillai over 2 years ago
Connect to SDC co-located server and check to see records from eye unidentifiable state to identify the issue.
Updated by Sreeshna T over 2 years ago
- Status changed from New to Feedback
- % Done changed from 0 to 100
count issue because of grade is good and error observation is eyetype unidentifiable considered as good in true positive case.
Updated by Sreeshna T over 2 years ago
- File error_observation.xlsx error_observation.xlsx added
found 6 patients whose gradability is bad and error observation is EyeTypeUnidentifiable from IDRSTL DB.
Updated by Thara S Pillai about 1 year ago
- Status changed from Feedback to Closed